Precision Digital Thermometer Thermocouple & RTD |
|

|
|
The INFT thermocouple meter offers very high
accuracy ensured by an advanced (patent applied for)
thermocouple linearizing system. Accuracy for most thermocouple
types is 0.2°C. The meter is front panel programmable
for any of 9 thermocouple types. The INFT RTD input
meter is programmable for any 6 to 6000ohm DIN or NIST
platinum RTD. It provides a precision ultra-low constant
current excitation to minimize self-heating errors and
maximize stability.
SPECIFICATIONS
Thermocouple Input: J, K, T, E, R, S, B, N, J
DIN
RTD Input: 100ohm Pt (385 or 392 curve) any 6
to 6Kohm NIST or DIN Pt and any linear RTD (10ohm Cu,
etc); 2, 3 or 4 wire
Step Response: 1 sec to 99.9%
Warmup to Rated Accuracy: 50 min
Operating Ambient: 0 to 50°C (32 to 122°F), 95%RH,
non-condensing
Storage Ambient: -40 to 85°C (-40 to 185°F)
Power: 115 or 230 Vac, 49-400 Hz; 10 to 32 Vdc
Power Consumption: 6 W nominal, 10 W max.
Normal Mode Rejection: 60 dB
Common Mode Rejection: 120 dB
Common Mode Voltage: 1500 V peak per Hv test
Resolution: 15-bit
Conversion: dual-slope technique
Reading Rate: 3/sec or 13/sec, 60 Hz; 3/sec or
12/sec, 50 Hz
Input Configuration: single-ended Polarity: unipolar/bipolar,
programmable
Span Adjustment: +0.00001 to 500,000, programmable
Offset Adjustment: 0 to 999,999 or 0 to -99,999;
programmable
Sensor Excitation: 24 V at 25 mA 160 mA for 100ohm
RTD 1.6 mA for 10ohm Cu
Display: red or green 6-digit, 14-segment, 13.7
mm (0.54"); 4 alarm indicators
Dimensions: 48 H x 96 W x 165 D mm (1.89" x 3.78"
x 6.5")
Panel Cutout: 45 H x 92 W mm (1.772" x 3.622");
1/8 DIN
Weight: 574 g (1.27 lb)
TTL Outputs: four, isolated open collector; rated
150 mA at 1 V sink, 30 V open
Dual Relays: form C, 7 A at 30 Vdc, 115 or 230
Vac
BCD Output: isolated, tri-state, TTL/CMOS compatible;
external 5 V supply for isolated; internal 5 V supply
for non-isolated
Four Relay Option: dual 7A relays and dual 1
A relays
Analog Output: 0-5 V/1-5 V/0-10 V/0-20 mA/4-20
mA, user selectable; 354 Vp isolation; 14-bit resolution;
0.1% accuracy, 50 msec step response
RS-232 Communications: 300/600/1200/2400/4800/9600/19.2k
baud; RJ11 4-wire connection; complete program setup
and message display capability; programmable to transmit
current display, alarm status, min / max, actual measured
input value and status, supports continuous or command
mode
RS-485 Communications: 300/600/1200/2400/4800/9600/19.2k
baud; RJ12 6-wire connection; addressable from 0 to
199
|
|
|
Input
Type
|
Range,
°F
|
Range,
°C
|
Range,
K
|
Accuracy*
|
|
|
Iron-Constantan
|
-346
to 1400°F
|
-210
to +760°C
|
63.2
to 1673.2K
|
0.2°C/0.3°F/0.2K
|
|
|
Chromel-Alumel
|
-454
to 2500°F
|
-270
to 1372°C
|
3.2
to 1645.2K
|
0.2°C/0.3°F/0.2K
|
|
|
Copper-Constantan
|
-454
to 752°F
|
-270
to 400°C
|
3.2
to 673.2K
|
0.2°C/0.3°F/0.2K
|
|
|
Chromel-Constantan
|
-454
to 1832°F
|
-270
to 1000°C
|
3.2
to 1273.2K
|
0.2°C/0.3°F/0.2K
|
|
|
Pt13%Rh-Pt
|
-58
to 3214°F
|
-50
to 1768°C
|
223.2
to 2041.2K
|
0.2°C/0.3°F/0.2K
|
|
|
Pt10%Rh-Pt
|
-58
to 3214°F
|
-50
to 1768°C
|
223.2
to 2041.2K
|
0.2°C/0.3°F/0.2K
|
|
|
Pt30%Rh-Pt6%Rh
|
+212
to 3300°F
|
+100
to 1820°C
|
373.2
to 2093.2K
|
0.3°C/0.5°F/0.3K
|
|
|
OMEGALLOY
*1
Nicrosil-Nisil
|
-454
to 2372°F
|
-270
to 1300°C
|
3.2
to 1573.2K
|
0.2°C/0.3°F/0.2K
|
|
|
Iron-Constantan
|
-328 to 1652°F
|
-200 to 900°C
|
73.2
to 1173.2K
|
0.6°C/1.0°F/0.6K
|
*1 - Registered Trademark of OMEGA ENGINEERING, INC.
*Accuracy (all±) includes maximum linearization error
|
Input
Type
|
Range,
°F
|
Range,
°C
|
Range,
K
|
Accuracy*
|
| RTD
1 - 10ohm Copper |
-328 to 392°F
|
-200
to 200°C
|
73.2
to 473.2K
|
1.0°C/2.0°F/1.0K
|
| RTD
2 - 0.00385 100ohm Pt |
-328 to 1652°F
|
-200
to 900°C
|
73.2
to 1173.2K
|
0.2°C/0.3°F/0.2K
|
| RTD
3 - 0.00392 100ohm Pt |
-328 to 1562°F
|
-200
to 850°C
|
73.2
to 1123.2K
|
0.2°C/0.3°F/0.2K
|
To Order (*insert
number code to complete model number)
|
Basic
Model
|
Power/
Display
|
Control
Output
|
Analog
Output
|
Serial
Output
|
Input
Signal
|
Description
|
|
INFT
|
(
* )
|
(
* )
|
(
* )
|
(
* )
|
(
* )
|
Temperature
panel meter |
|
|
(
* )
|
(
* )
|
(
* )
|
(
* )
|
(
* )
|
Split
meter with remote display |
| |
0
|
|
|
|
|
115
Vac power, red LED display |
| |
1
|
|
|
|
|
230
Vac power, red LED display |
| |
2
|
|
|
|
|
115
Vac power, green LED display |
| |
3
|
|
|
|
|
230
Vac power, green LED display |
| |
4
|
|
|
|
|
10-32
Vdc power, red LED display** |
| |
5
|
|
|
|
|
10-32
Vdc power, green LED display** |
| |
|
0
|
|
|
|
Four
NPN open collector transistors |
| |
|
1
|
|
|
|
Isolated
parallel BCD |
| |
|
2
|
|
|
|
Two
7 A relays |
| |
|
3
|
|
|
|
Two
7 A relays and two 1 A relays** |
| |
|
|
0
|
|
|
No
analog output |
| |
|
|
1
|
|
|
Isolated
analog output |
| |
|
|
|
0
|
|
No
serial output |
| |
|
|
|
1
|
|
Isolated
RS-232 |
| |
|
|
|
2
|
|
Isolated
RS-485 |
| |
|
|
|
|
(
* )
|
Specify
range signal from chart below.* |
** 4 Relay option is not available with 10-32 Vdc Power Option
*Specify C (°Celsius), F (°Fahrenheit), or K (Kelvin)
|
|
|
† All dollar amounts on this site are shown in US currency.
|
|
 |
 |
 |
 |
 |
 |
 |
 |
 |
Part Number Builder |
 |
 |
 |
 |
 |
 |
|
|
|
Option Descriptions
|
|
(1)
|
Meter Type
|
| |
Select INFT for INFINITY® TEMPERATURE input meter INFZT for TEMPERATURE input split meter with remote display
|
(2)
|
Power and LED Color
|
| |
Select 0 for Designates 115 Vac power and red LED display 1 for Designates 230 Vac power and red LED display 2 for Designates 115 Vac power and green LED display 3 for Designates 230 Vac power and green LED display 4 for Designates 10-32 Vdc power and red LED display 5 for Designates 10-32 Vdc power and green LED display
|
(3)
|
BCD and Control Outputs
|
| |
Select 0 for Four optically isolated open-collector outputs 1 for Isolated parallel BCD output (BCD1) 2 for Isolated dual 7 A relays (REL1) 3 for Isolated dual 7 amp and dual 1 amp relays (REL4), not available with DC power
|
(4)
|
Analog Output
|
| |
Select 0 for No analog output 1 for Isolated analog output (AN03)
|
(5)
|
Serial Outputs
|
| |
Select 0 for No serial output 1 for Isolated RS-232 serial output (RS24) 2 for Isolated RS-485 serial addressable output (RS24)
|
(6)
|
Input Options
|
| |
Select RTD1 for 10 ohm Copper RTD RTD2 for 100 ohm Pt RTD with 0.00385 alpha RTD3 for 100 ohm Pt RTD with 0.00392 alpha J for Iron-Constantan K for Chromel-Alumel E for Chromel-Constantan T for Copper-Constantan B for Pt30%Rh-Pt6%Rh S for Pt10%Rh-Pt N for Nicrosil-Nisil DIN J for Iron-Constantan
|
(7)
|
Temperature Units (Thermocouple/RTD configurations)
|
| |
Select C for °C (Celsius) F for °F (Fahrenheit) K for Kelvin
|
| NOTE: All combinations may not be valid, check spec sheet for valid part numbers. |
| | | | | |
|
 |
 |
|
|
|
|
Keywords: INFINITY® Precision Digital Thermometer Thermocouple & RTD
|
Back to top |
|
|